Main Features:
Significantly improved ECU performance and innovation.
Clarity in results with the 3rd Generation Integrated Multi Axis Control System (IMACS2).
Fully integrated and compact in design, a perfect solution for both static and mobile testing facilities.
The new Hydraulic Power Supply (HPS) uses inverter technology coupled with a high performance labyrinth bearing actuator to deliver 19 kN of force.
All-in-One computer control with the latest All-in-One touchscreen PC technology.
Easy-to-use and ergonomic testing system.
Interchangeable transducers and load cells with "plug and play" signal conditioners allowing for quick and easy transition of test set-ups.
Optional air compressor that comes integrated.
Technical Specifications
Load Capacity: Static: 19kN, Dynamic 17kN
Actuator Type: Labyrinth Bearing
Frequency Range: 0.01 to 70Hz sinusoidal loading
Actuator Stroke: 30mm (+/- 15mm stroke)
Specimen Size (dia xh): 100 x 150 mm nominally, 50 x 135 mm, 38 or 50 x 110 mm
Cell Dimensions (dia xh): 270 x 390mm
Confining Pressure: 0 to 225 kPa
Air Compressor and Dryer: Low noise, integrated, automated operate-on-demand (optional)
Easily interchangeable load cells test
Noise Level: Less than 70db at 2m
Computer Control: Integrated all-in-one PC touchscreen (optional)
Supplied complete with high quality pre-filtered oil for longer life
Temperature Range: -10 ° C to + 70 ° C *
Cooling Rate: Typically cools to + 4 ° C in under 30 minutes (at an ambient temperature of + 23 ° C)
Temperature Accuracy: +/- 0.2 ° C (at temperature probe placed close to the specimen)
Dimensions (hxwxd): 1,359 x 1,350 x 739mm
Weight: 275kg (including oil)
Transducers
Load Cell: Low profile pancake type
Pressure: 0 to 225 kPa
Built-in LVDT Actuator: 30mm Stroke
Temperature Probe Range: from -25 up to +80° C
Plug-and-Play: Up to 4 interchangeable on-specimen displacement transducers, plus easily interchangeable load cells
On-Specimen Displacement: 3 clips on +/- 0.5mm LVDTs Compatible with up to 4 (Various optional loose-core, strain gauge transducers available)
Services
Power (with air compressor): 220V–240V, 50Hz, single phase, 22A 208V, 60Hz, single phase, 25A
Power (without air compressor): 220V–240V, 50Hz, single phase, 17A 208V, 60Hz, single phase, 20A
Hydraulic Oil: Pre-filled with high specification pre-filtered, ISO 46 Premium Mineral Oil
Air: Clean dry air at 450-800kPa; 2 L/sec (Optional integrated air compressor available)
Control & Data Acquisition—IMACS2
Configuration: Fully integrated
Real Time Digital Computer Control: 32-bit Processing
Acquisition Speeds: 200kHz (simultaneous on all channels)
Data Over-Sampling: At least 4x
Analogue Outputs: 8 BNC connectors for raw data logging
Control: 2 axis control (actuator and confining pressure)
Data Resolution: 24-bit auto-ranging data acquisition
Communication: Ethernet at 10/100Mb/s, Ethernet at 1Gb/s
Analogue Inputs: Auto-calibrate on power up
Acquisition: Up to 8 Channel data acquisition (actuator displacement, axial load, 3 to 4 on-specimen displacement transducers, confining pressure and temperature)